11.2.21 E : 31 – Warm-up timer
The sensor goes through a stabilization period when power is first applied. This option allows the user to adjust the time that is
applied before readings are displayed / transmitted.
From the menu system select menu option: E: 31 and press ENTER.
Using the INCREASE and DECREASE buttons, set the required warm-up time.
Press ENTER to store the new value.
Note: Pressing the MENU button rather than the ENTER button exits without any change.
11.2.22 E : 47 – Analogue Output FSD
This menu option allows the full-scale gas level to be selected that relates to the 4 to 20 mA output.
From the menu system select menu option: E: 47 and press ENTER.
Using the INCREASE and DECREASE buttons, adjust the FSD to the required level.
Press ENTER, “----“ will be displays to indicate that the new setting has been accepted.
Note: Pressing MENU instead of ENTER will exit the menu without changing the FSD.
Press the MENU button to close the menu system
Normally warm time is 45 seconds.