Enmet EN Series Manuale Utente Pagina 10

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 19
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 9
Sensor/Transmitter, Toxic Gas & Oxygen ENMET Corporation
8
Appendix A: Interference Table
For Selected Sensors
Ammonia (3E-100 sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
ethanol
1000
1
isopropanol
450
1
hydrogen sulfide
14
10
methanol
1200
3
hydrogen
1000
80
carbon monoxide
300
100
sulfur dioxide
25
-
21
nitrogen dioxide
50
25
hydrogen cyanide
10
-
18
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
chlorine
5
unsaturated hc (ethylene)
1.0 %
saturated hydrocarbons
abundant
Carbon Monoxide (3E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
hydrogen
1000
450
nitric oxide
100
25
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
ammonia
100
chlorine
5
hydrogen cyanide
10
ethylene
2.0 % *
carbon dioxide
5,000
methane
10,000
sulfur dioxide
10 *
hydrogen sulfide
10 *
nitrogen dioxide
10 *
isopropanol
1,025 *
gasoline vapor
saturated *
* With onboard filter; continuous high level exposure may reduce the filter efficiency
Vedere la pagina 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 ... 18 19

Commenti su questo manuale

Nessun commento